Product category:
Mechanical testing equipment
News Release from: Admet | Subject: Precise Digital
Edited by the Laboratorytalk Editorial
Team on 07 November 2005
Resurrection of Baldwin universal
testing machine
Digital testing control system, new motor, and complete rewiring brings 1950s-vintage 200,000lb (90 tonne) testing machine back to life at University of Miami
The University of Miami College of Engineering's civil, architectural, and environmental department has several state-of-the-art measurement and testing tools that are used in education and research Until recently, one of its largest testing units was also the least useful
This article was originally published on Laboratorytalk on 26 May 2003 at 8.00am (UK)
Related stories
Digital controller for materials testing
Full-function, self-contained test controller for new and retrofit applications is designed for both manually operated and servo-equipped electromechanical and hydraulic testing machines
Managing concrete evidence
Software manages the entire concrete testing and certification process with automatic generation of compressive strength reports and posting of results to private web portals
A 90 tonne Baldwin screw gear universal testing machine was not functioning.
Installed in 1959, the analogue machine fell into disrepair after many years of service.
Repairs were considered, but the company had long since disappeared.
Further reading
Retrofit universal testing machines to save cash
Digital controllers and Windows-based software update electro-mechanical and hydraulic load frames from many manufacturers to meet new requirements and certification standards
Innovative tabletop universal testing machine
Features removable controller, horizontal and vertical operation, protection against dust/liquids and various load-speed capacities
Tenacity meter for textile testing
The unit can be retrofitted to existing testing machines to eliminate potential errors when interpreting analogue dials, grid paper and hand-writing results to paper
As with many older universal machines, the test frame itself was fine, it was the motor, controllers and, in this case, an old analogue vacuum tube console, that had failed.
Professor Ronald Zollo sought to make the machine operable in as cost effective manner as possible.
A few years ago, working with Admet on an unrelated project, he found his opportunity.
The University of Miami College of Engineering's civil, architectural and environmental engineering department in Coral Gables, Florida, prepares students for careers in building structures, highway, pipeline, bridge and other infrastructure design and construction.
For many students, lab work is an integral part of their education.
The university makes a range of engineering and testing equipment available to students for course and project work.
In addition, advanced degree students and faculty conduct innovative research into highway, bridge, building, infrastructure and other systems.
The lab performs American Society for Testing and Materials (ASTM) testing on numerous materials and structures, including beams, pipes and reinforced concrete.
Funded research helps support faculty as well as graduate and undergraduate laboratory activities.
In today's world, student laboratory experience is often an adjunct to computer modelling.
Students conduct tests to verify computer results to see how the data correlates with actual physical tests.
The centerpiece of the lab is a large Baldwin universal testing machine.
The machine, installed around 1959, is a 90 tonne four-column, screw-type test frame.
It is mounted into the floor with sub-floor works and is surrounded by a structural floor system.
Although state-of-the-art at time of installation, the times have changed.
In its original configuration, a vacuum tube-equipped console controlled the crosshead position and collected some rudimentary readings.
The 1m3 refrigerator-sized analogue control console occupied valuable laboratory space.
Remarkably, the machine continued to operate until the mid 1990s.
Commented Professor Zollo: "I've been working to restore the lab to fully operational condition so I was quite disturbed that this machine, which is a centerpiece in the laboratory, was not functional when I knew that we could use it".
Over the years it had become impossible to find replacement parts for the console as technology evolved to digital controls and computer-based programs.
"The machine was essentially inoperative.
"We could move the crosshead up and down and use it in a very limited sense, but it wasn't really functional," commented Zollo.
In addition to limiting their ability to test large structures, the students were denied the opportunity to work with equipment that they would encounter in an industrial setting after completing their education.
Finding a solution.
At $250,000 a new machine was out of the question.
So, the Baldwin sat unused for many years until Zollo found the solution.
He discovered Admet as a result of his own research activities.
For years he has been studying fibre reinforced concrete.
Through his association with others in the American Concrete Institute he became aware of a company in Tennessee which had used Admet equipment.
When he visited its plant he noticed an older universal testing machine that had been retrofitted by Admet.
Not sure if the same type of upgrade could be performed on the Baldwin, Zollo invited an Admet engineer to visit the university.
The engineer evaluated the machine and determined that it could be retrofitted with digital controls.
The estimate was $18,000 for a full retrofit that included installing a new Admet MTestWindows personal computer-based digital testing software system and replacing the motor.
The machine had to be entirely rewired and a new connector/controller box added under the floor level.
Digital revival.
Now, an Admet Precise Digital controller measuring 25x30x15cm has made the machine fully operational.
In fact, it's better than new since the Precise Digital not only controls the crosshead speed and position, it acquires data on speed, displacement, peak load, etc, and feeds that back into the MTestWindows program.
Lab technicians can attach separate load cells, as well as strain gauges and linear variable differential transformers (LVDTs), to the Admet system.
"Anything that can create an analogue signal can be converted to digital on the Admet system and then put into the computer for plotting," said Zollo.
Student test data is collected on the PC and made available over the university's internet network.
Although the Admet systems will perform many of the calculations, the students are expected to use a spreadsheet or other program to calculate and interpret the data.
"We want them to see what is done so they can understand how a program can be written to do the calculations," explained Zollo.
"It's great that we were able to bring this machine back to life so inexpensively.
"It shows the progress that's been made when a cubic foot box with state-of-the-art electronics can replace a refrigerator-sized console," concluded Zollo.
• Admet: contact details and other news
• Email this article to a colleague
• Register for the free Laboratorytalk email newsletter
• Laboratorytalk Home Page