News Release from: Artel | Subject: Extreme Pipetting Expedition
Edited by the Laboratorytalk Editorial
Team on 13 November 2007
Extreme pipetting: dry heat causes 35%
error
At Death Valley for mission number 3 of the Extreme Pipetting Expedition, Artel found that pipettes under-deliver by up to 35 percent in dry and hot environments
While volume delivery errors were partially reduced by pre-wetting pipette tips, under-delivery still persisted, and the pipettes were found to operate out of specification in most instances Laboratory technicians commonly encounter dry heat in their own work environments, which, although not as extreme as Death Valley, can nonetheless affect pipette performance and compromise assay results
This article was originally published on Laboratorytalk on 12 Feb 2007 at 8.00am (UK)
Related stories
Artel to launch extreme pipetting expedition
At Pittcon 2007, Artel will announce its Extreme Pipetting Expedition, a multi-phase, year-long scientific study to demonstrate the impact of environmental conditions on laboratory data integrity
Pipetting expedition highlights sources of error
During this multi-phase, year-long scientific study, Artel will visit locations with extreme ranges of common laboratory conditions, such as high and low temperature, humidity and barometric pressure
For example, laboratories using analytical instruments, ovens, incubators, freezers and other devices requiring high power or open flames are prone to heat and low humidity.
Heating and air-conditioning systems not equipped with humidifiers may also contribute to dry heat.
To ensure pipetting accuracy and precision and strengthen data integrity, it is essential that laboratory scientists understand and compensate for this source of error.
Further reading
Extreme pipetting challenge takes on Yellowstone
At Yellowstone, Artel will test how pipetting liquids at a temperature different than the pipette (thermal disequilibrium) affects delivered volume and data accuracy and precision
Liquid temperatures create large pipetting errors
Thermal disequilibrium is prevalent in today's laboratories, as many assays and tests require the addition of liquids that are warmer or colder than the laboratory environment and the actual pipettes
Streamlining forensic laboratory quality assurance
PCSplus provides Nist-traceable pipette performance verification in minutes, strengthening accuracy and precision in DNA testing, toxicology screening and other critical forensic laboratory processes.
"Laboratories often experience changes in humidity and temperature throughout the course of a year and data may be at risk," says George Rodrigues, senior scientific manager, Artel.
"Humidity and temperature can also vary across laboratories in different regions, between laboratories on different floors, and even in different sections of the same laboratory".
Like in previous missions, error caused by the environmental conditions at Death Valley was greater when pipettes were set to their minimum volumes than when they were set to their maximum volumes.
A two-microlitre pipette set to deliver its minimum volume of 0.2microliter without pre-wetting under-delivered by 34.7 percent.
Pre-wetting reduced the error to 30.9 percent, but the pipette was still operating outside its minimum-volume accuracy specification of 12 percent.
When set to its maximum volume of two microlitres without pre-wetting, the pipette under-delivered by 7.0 percent without pre-wetting and by 4.8 percent with pre-wetting, both outside of the manufacturer's accuracy specification.
When working with larger liquid volumes, errors were also induced by the hot and dry conditions, but on a smaller magnitude.
Without pre-wetting, a 20-microlitre pipette at its minimum volume under-delivered by 16.8 percent, which is a significantly lower inaccuracy than exhibited by a non-pre-wet two-microliter pipette at its minimum volume (34.7 percent error).
Pre-wetting the 20-microlitre pipette prior to dispensing its minimum volume of two microlitres reduced error to 7.4 percent.
"Based on data taken at Death Valley, we highly recommend pre-wetting pipette tips to reduce volume loss when pipetting in dry and hot environments," says Doreen Rumery, laboratory technical manager and quality control manager, Artel.
Liquid handling error experienced in dry heat is largely due to evaporation.
The evaporation of a miniscule amount of liquid inside the pipette tip has a large effect on pipetted volumes, especially when target volumes are in the microliter range.
When one microlitre of liquid evaporates, it converts into more than 1000 microlitres of gas, expanding by a factor of 1250 to 1450 depending on temperature.
This expansion prevents the pipette from aspirating the desired target volume.
Death Valley was selected as the site for Mission 3 because of its iconic hot and dry climate.
Surrounded by mountains, the valley captures and retains extreme amounts of heat to create one of the hottest environments on earth, commonly exceeding 49C.
Death Valley is also the driest locale in the United States, receiving fewer than 50mm of rain annually.
The Extreme Pipetting Expedition is a multi-phase, year-long scientific study to illustrate the impact of laboratory environmental conditions on pipetting performance and data integrity.
Previous missions explored how barometric pressure and how pipetting warm and cold liquids affect data integrity.

Not your typical laboratory environment: pipetting in Death Valley USA
.