What's that noise?
Keith Hall of LabLogic Systems tracks down the source of background during counting of low-level radioactivity in RHPLC eluates, and describes an effective way of keeping it to a minimum
Among radiochromatographers, the terms 'noise' and 'background' are used interchangeably to refer to the radiation recorded by the detector that is extraneous to the samples under investigation and must therefore be eliminated from calculations Talk of 'background' is misleading because it implies that this detected but irrelevant radiation originates from sources distant from the focus of investigation, such as other equipment operating in the laboratory, but in fact this is far from the truth
This article was originally published on Laboratorytalk on 2 Jul 2008 at 8.00am (UK)
Related stories
Scintillation cocktail preferred for radio-HPLC
A leading CRO recently secured a contract in which the sponsor required the use of Flowlogic 1:1 scintillation cocktail for all radio-HPLC work
Imaging for faster radio-synthesis purity checks
AR 2000 imaging scanner from LabLogic Systems gives radio-synthesis laboratories a fast and accurate means of measuring the purity of their product in real time without the need for a full HPLC system
Noise arising from electronic equipment - even from a single photo multiplier tube in a radio flow detector - is so low as to be unmeasurable; and radioactivity present in an ambient state or from the action of cosmic rays does not register much higher up the scale.
The emissions of radio isotopes are measurable of course, but only strong types such as 40K gamma are detectable at any distance; tritium, one of the most commonly used isotopes in ADME studies, travels less than 6mm in air.
Paradoxically, almost all the 'background' measured by the radio flow detector in a radio HPLC (RHPLC) system originates in the foreground and very close to home, because it is generated by the detector's own photo multiplier tubes (PMTs).
To distinguish valid data from noise, the detector has two PMTs which must respond to the same beam of light within 20 nano-seconds for the signal to be recognised as valid.
Logically, it is also possible for both tubes to respond in the same time frame but to different events (a phenomenon described as cross-talk), and it is this that accounts for 90 per cent of what we call noise or (very inaccurately, as we can now see) background.
RHPLC detector manufacturers use various means of excluding cross-talk to reduce background, which in turn ultimately improves limits of detection (LOD).
The most successful in this respect (and the most sensitive) is the SS-Ram from IN/US Systems, which has photomultiplier tubes with a diameter of just 31.75mm (1.25inch) to maximise signal-to-noise ratios.
By blanking off a proportion of the PMT phases from each other, this arrangement achieves ultra-low backgrounds - typically <4cpm for 3H and <1cpm for 14C, with corresponding counting efficiencies of > 60% and > 90% respectively. Request a free brochure from Lablogic Systems ...