Product category:
Heating: ovens, furnaces, autoclaves
News Release from: LTE Scientific | Subject: Steri-Master
Edited by the Laboratorytalk Editorial
Team on 28 May 2002
Programmable control systems fit the
bill
Multi-programmable microprocessor autoclave control system is said to be one of the most flexible, safe, and user-friendly available today
The Steri-Master makes the selection and setting of sterilisation cycle parameters easier than ever and operates with a wide variety of load types Holding up to eight cycle types in its memory, the Steri-Master has a data printer and RS232 output as standard, which makes it ideal for laboratories requiring accurate records to comply with regulatory requirements or quality control standards, such as ISO9000 or GLP
This article was originally published on Laboratorytalk on 19 Oct 2001 at 8.00am (UK)
Related stories
New ovens and incubators
Designed to meet the increasing demands of today's laboratory and process facility, the brand new OP ovens and IP incubators series from LTE Scientific offers maximum flexibility and performance
Analytica preview: LTE Scientific
Manufacture of autoclaves, incubators, ovens, and freeze dryers will be featuring the latest additions to its range, including new high capacity sliding door steam sterilisers
The new system, which complements LTE's Uni-program single cycle model for repetitive applications, is supplied as standard on all Kestrel and CV Series autoclaves.
For Falcon and Osprey sterilisers, customers can specify either the Steri-Master or the uni-program control system.
Providing accurate cycle reproducibility and traceability, parameters for each cycle are entered via a touch-pad with commands, prompts and status messages shown clearly on an LCD screen.
Further reading
Interphex preview: LTE Scientific
Expanded range of autoclaves, incubators, ovens, freeze-dryers, and environmental rooms on display, including brand new laboratory ovens and microbiological incubators
Autoclaves of every size
Available in sizes from 300 to 7000 litres, and featuring a two-stage air removal phase, this value-for-money range offers great flexibility
Many advanced features have been incorporated, including alpha-numeric inputting for programming users' own cycle titles, and a discard cycle option - for reducing cycle times for waste loads.
To provide additional safety in use, an overall cycle time option is also included, offering additional protection to users processing loads with validated cycle times.
A 'time-out' function ensures the autoclave door cannot be opened until the load has cooled below 80C.
For optimum security, three levels of access are built into the system - Operator, Supervisor, or Engineer.
With restricted access codes, the Supervisor mode is used to set or alter program parameters while the Engineer mode enables only qualified personnel to adjust the machine settings.
The Steri-Master system is equipped with a sophisticated 'watchdog' facility, which monitors the predetermined temperature limits on all three channels (the drain, the load, and the load simulator), as well as other aspects of the autoclave1s status.
Error messages are displayed and printed, covering a wide range of events, while other fault messages are designed as an aid to trouble-shooting and maintenance.
In addition to the parameters available on the standard autoclave range, vacuum models include purge type (vac/non-vac), number of pulses, final cool method, positive steam purge option, pre-cool method and drying method.
The control system on all vacuum models has an automatic leak test cycle.
The data printer on vacuum models also prints relevant vacuum and pressure values.