Water determination in polyester powder
The water, which is very strongly bound in the plastic, is thermally driven off from the sample and transferred to the KF cell with a carrier gas stream
Most plastics dissolve only slowly or not at all in organic solvents This is why Metrohm recommends the heating method for such applications
This article was originally published on Laboratorytalk on 2 Jul 2008 at 8.00am (UK)
Related stories
Analysis of mercury in waters using voltammetry
Jonathan Bruce, Applications Manager for Metrohm UK, provides this comprehensive introduction to analysis of mercury in waters
The water, which is very strongly bound in the plastic, is thermally driven off from the sample and transferred to the KF cell with a carrier gas stream.
Because of the very low water content the determination with the 831 KF Coulometer or 756 KF Coulometer is to be preferred.
The heating temperature should be optimized by applying a temperature gradient between 50-250C with the 774 oven sample processor.
This preliminary experiment shows that the water is released at low temperature (approx 90C), but also that at higher temperatures no side reactions take place.
For this reason a heating temperature of 220C was used to decrease the determination time.
For the determinations between 0.3 and 3g of sample were weighed in a vial and hermetically sealed with a septum cap.
Six determinations were run at 220C with the following result: water content 0.50% +/-0.01%. Request a free brochure from Metrohm UK ...