Visit the Metrohm UK web site
Click on the advert above to visit the company web site

Product category: Chromatographs: liquid, ion, gel, HPLC
News Release from: Metrohm UK | Subject: Primus standards
Edited by the Laboratorytalk Editorial Team on 18 December 2007

Ion chromatography standard solutions
are certifed

Each Metrohm ion chromatograph is delivered together with a Primus standard (primary multi-ion standard; multi-ion standard for the determination of anions or cations) from Fluka

Quantitative determinations in ion chromatography are based on calibration methods Reliable calibration curves can only be obtained by using calibration standards that are of perfect quality

Primus standards are not only used for calibration, but also for demonstrating that the instruments are functioning properly after installation.

Primus standards are prepared from characterised primary materials (Li2CO3, KCl, CaCO3, MgO, NaCl, NaF, NaBr, NaNO3, Na2HPO4, Na2SO4) and come in 50ml bottles.

The anion multi-ion standard contains 10ppm each of fluoride, chloride, bromide, nitrate, phosphate and sulphate and can be reordered from Fluka (order number: 89886).

The cation multi-ion standard contains 10ppm each of lithium, sodium, potassium, calcium and magnesium.

It can also be reordered directly from Fluka (order number 89316).

The solutions are supplied in HDPE bottles contained in a sealed aluminum bag.

In this way, both solutions can be stored in their original packaging for up to five years.

The manufacturer guarantees that the concentrations will remain unchanged throughout this period.

In addition to the Primus standards, there are other standard solutions that can be used for the validation of ion chromatographs.

For the operational qualification (OQ) of systems of the type 850 Professional IC, a Metrohm OQ Set has been defined (order number 6.5333.000) which, in addition to the appropriate documentation and methods, contains a back-pressure capillary and a suitable Teflon capillary as well as a standard set of six solutions.

These standard solutions contain bromide at concentrations of 5, 10, 20, 50, 100 and 1000ppm.

They can be used for the validation of systems with conductivity, UV/vis or electrochemical detectors.

The six solutions in the standard set are certified and are produced by Fluka.

Like the Primus standards, they can also be reordered from Fluka (order number 12674). Request a free brochure from Metrohm UK ...

Metrohm UK: contact details and other news