Product category:
Laboratory and scientific consultancy services
News Release from: Sira
Edited by the Laboratorytalk Editorial
Team on 30 October 2002
Improved pressure balance calibration
service
Specialist in calibration, testing, and certification is now able to offer further reduced measurement uncertainties under its gold service for pressure balances
Sira Test and Certification offers three classes of accuracy in pressure balance calibration, provided by staff with an in-depth knowledge of industrial calibration and instrumentation 35-50ppm gold service (mass and effective area) involves weighing the piston and each weight to an uncertainty of approximately 8ppm (or less if required)
This article was originally published on Laboratorytalk on 9 Aug 2004 at 8.00am (UK)
Related stories
Rose-tinted glasses
Remote optical sensing evaluation (Rose) project enables company to push boundaries and maintain its reputation as a dynamic force in the field of environmental monitoring
Time is running out to avoid an explosive future
Off-the-shelf packages to help the self-employed, small and medium-sized companies comply with Dangerous Substances and Explosive Atmosphere Regulations
The effective area of the piston assembly is determined by direct comparison with a pressure balance, calibrated at a National Standards laboratory, yielding an uncertainty of the effective area measurement which is usually less than 50ppm.
Typically, the calibrated range of the instrument is 90-95% of the working range.
100ppm silver service (standard service) calibrates instruments in a similar way to the 50ppm mass and effective area service.
The uncertainty of the principal measured pressures is around 100ppm (up to 20,000 psi) increasing to 250ppm (up to 60,000 psi).
200ppm bronze service (five point calibration); the pressure generated by the pressure balance is measured at five points across each range with an uncertainty of around 200ppm.
Sira says it is able to calibrate all makes of equipment, including dual and single piston oil operated pressure balances up to 4100bar.
Sira can also calibrate single, dual and triple piston gas operated pressure balances up to 400bar.
Sira also offers repair services as well as training courses in several areas of calibration.
• Sira: contact details and other news
• Email this article to a colleague
• Register for the free Laboratorytalk email newsletter
• Laboratorytalk Home Page