Team on 20 August 2002
Free fatty acids analysis breakthough
A novel, easy to use thermometric titration method permits rapid, accurate, automated analysis of the free fatty acid (FFA) content of edible fats and oils with unprecedented precision and accuracy
The free fatty acid (FFA) content of edible fats and oils is an important indicator of rancidity It is used by producers to control the quality of their product, and by commercial users such as manufacturers of fried snack and convenience foods to control frying processes
This article was originally published on Laboratorytalk on 19 Aug 2002 at 8.00am (UK)
Related stories
Analysis of moisture in oily, greasy matrices
The incompatibility of the highly polar Karl Fischer reagent with non-polar samples can introduce difficulties in the determination of moisture by this traditional approach
Analysis of acid mixtures
Method chaining feature involving two separate titrations solves the problem of analysis of a mixture of nitric, acetic, and phosphoric acids
Regular monitoring of frying oils is used to control the quality of the product as well as determine replenishment and replacement schedules.
The official AOCS method involves the titration of an ethanolic solution of the fat or oil with ethanolic potassium hydroxide to a visually-determined phenolphthalein endpoint.
Titrations employing visual indicators require the analyst to be able to distinguish subtle differences in shade and colour, and to be well trained in the technique.
Further reading
Surfactant analysis 'not all froth and bubble'
Thermometric titration technology can make surfactant analysis easier; a simple, long-lasting thermometric probe is the easy alternative to toxic reagents or expensive electrodes
Solving an explosive problem
Analysis of acid mixtures in explosives is far from straightforward with conventional titration, but some cleverness reduces the complexity enormously
A Fab answer
Solving problems in the analysis of solutions used in the semi-conductor fabrication industry, using modern titration technology instead of complex manual methods using visual endpoint indication
The task may be rendered more difficult if the color or clarity of the sample solution varies.
Differences in analytical results between analysts are common.
Automated potentiometric titration methods have been developed, with a pH electrode used to indicate the titration endpoint.
The operation of glass membrane pH electrodes in an oily, ethanolic solution requires careful monitoring, as the surface of the electrode becomes both dehydrated and fouled by the sample solution.
The junction of the reference electrode may also become fouled by the oily sample matrix.
To restore satisfactory performance, it is necessary to frequently re-hydrate the glass membrane by immersion in water.
The analysis of poorly dissociated, weakly acidic species in a non-aqueous environment is not an ideal application for potentiometric titrimetry.
There is a clear need for a robust, accurate and precise automated titrimetric procedure which is substantially independent of operator technique, and where the sensor is unaffected by the nature of the sample.
Thermometric titrimetry has proved to be a robust, reliable technique for a wide variety of analytes across many industries.
A simple thermistor is used to sense the change in solution temperature which signals the titration endpoint.
Since only the change in temperature is of interest, there is no requirement to calibrate the sensor.
Unlike potentiometric probes, the thermistor may be regarded as a 'zero maintenance' sensor.
A modern thermometric titration system employs powerful algorithms to optimally condition the temperature signal and permit the computation of derivatives to accurately locate endpoints.
A requirement of many thermometric titration applications is that the temperature change at the endpoint is sufficiently great to overcome other thermal effects such as heats of mixing and loss of heat from the system.
This is often overcome by using relatively strong titrants, and commensurately higher sample amounts.
In the case of the determination of weak acids in non-aqueous media, the amounts present are relatively low and it is necessary to use a low strength titrant in order to obtain meaningful titres.
Such applications might be considered to be antipathetic to the application of thermometric titrimetry.
However, Vaughan and Swithenbank discovered that it was possible to obtain highly accurate results for a range of weakly acidic species by a technique which could be described as catalysed endpoint thermometric titrimetry, or CETT.
In the determination of weakly acidic species, the sample was dissolved in anhydrous acetone and titrated with KOH dissolved in an alcohol.
After all acidic species have been reacted with the titrant, the first trace of free hydroxyl ion catalyses the exothermic condensation of acetone to diacetone alcohol.
Multitrator has found this to be a satisfactory and practical technique for the determination of phenols in coal tar distillate (the purpose for which it was originally developed), but less satisfactory for the routine determination of FFA in edible fats and oils.
Multitrator has discovered that another catalysed endpoint solvent system gives much improved analytical precision.
The base-catalysed reaction of acetone and chloroform is well known to be strongly exothermic.
In the context of the thermometric titration of weak acids, this reaction has been found to be safe and reliable.